Gas sampling and gas conditioning technology 2-1.1.7-ME
16.3.1 CHANGE AND CLEANING OF THE BYPASS INJECTOR (180°C VERSION)
For cleaning the injector of the 180°C-vesion it is not necessary to remove it
out of the Bypass-T. Clean the nozzle mechanically or put the complete
cross into an ultrasonic bath. For the 320°C-version do not clean the com-
plete cross in an ultrasonic bath because the graphite packing could be de-
stroyed.
A cleaning with compressed air has to be carried out from the sample outlet
side (jet pipe).
We recommend the following steps for the change of the bypass injector:
Unscrew the jet pipe (pos. 2) from the Bypass-T;
Push out the injector nozzle from the opposite side (tool pos. 3, Fig. 18) and change it with
sealing;
Push the injector into the T up to the block and press it tightly. Use the mounting tool.
16.3.2 CHANGE OF THE O-RING AT THE BYPASS INJECTOR (180°C-VERSION)
Figure 18 shows the injector in the 180°C-version.
Injector complete, SP2000-H/DIL
Figure 18 Bypass Injector unit with o-ring ( 180°C-Version )
We recommend the following steps:
Unscrew the jet pipe (pos. 2) from the Bypass-T;
Push out the injector nozzle from the opposite side (tool pos. 3, Fig. 18);
Remove the thrust collar and the o-ring;
Push new o-ring on the nozzle;
Put on the thrust collar;
Push the injector into the T up to the block and press it tightly. Use the mounting tool.