22
Gas sampling and gas conditioning technology 2-1.1.7-ME
An extract from an injector data sheet is shown below.
Injektor-Datenblatt / Injector data sheet
Betriebsdruck
Operating pres-
sure
[bar]
Unterdruck ohne kritische Düse
Vacuum without critical orifice
[bar]
Unterdruck mit kritischer Düse
Vacuum with critical orifice 5,1l/h
[bar]
Überprüfung des Verdünnungsfaktors
Check of the dilution ratio
Messgasdruck atmosphärisch
Sample gas pressure atmospheric
Kritische Düse
Critical nozzle
Verdünnungsgasdruck
Dilution gas pressure
Messwert d. verd. Gases
Meas. value of the dil.
gas
Figure 9 Extract from an injector data sheet
For the operation of an installed bypass injector, the necessary pressure must be set on the pressure
regulator provided (on left next to precision pressure regulator).
The gas pressure must be higher than 0.7 bar, as the vacuum valve
mounted on the probe has an opening pressure of 0. 7 bar.
When using the control panel, the respective ball valve for the test
gas supply must be opened and the flow rate set on the flow meter.
Betriebsdruck
Operating pressure
[bar]
Ansaugvolumenstrom bei 1bar
abs. Suction flow at 1bar abs.
[l/h]
Ansaugvolumenstrom bei 0,9bar abs.
Suction flow at 0,9bar abs.
[l/h]
Figure 10 Bypass injector data sheet