6.8
Technical specications and illustrations are without
obligation, subject to modications. 09.96/06.06
Pre-cooling unit version VC-... VC-1 VC-1-SL VC-2-L VC-2-SL
Part No. 230V 50-60Hz 03K1000 03K3000 03K4000 03K5000
Part No. 115V 50-60Hz 03K1000 03K3000a 03K4000a 03K5000a
Jet-Stream heat exchanger out of Duran glass EC-G 1x 1x 2x 2x
Forced ventilation no yes yes yes
Autom. condensate removal with SR25.1 1x* 2x*
Sample gas connection 2x GL18-6 2x GL18-6 4x GL18-6 4x GL18-6
Condensate connection 1x GL25-12 1x DN4/6 2x GL25-12 2x DN4/6
Power consumption 25 VA 20 VA 30 VA
Gas flow rate recommended, (other flow rates possible) 1x 250 Nl/hr 1x 250 Nl/hr 2x 250 Nl/hr 2x 250 Nl/hr
Weight approximately 3,5 kg 7,5 kg 9 kg 11 kg
Sample gas pressure max. 3 bar g
∆P per heat exchanger 4 mbar at 300 Nl/hr
Stagnant space per heat exchanger approx. 70 ml
Sample inlet temperature max. 180 °C
Ambient temperature +2 °C to +45 °C
Storage temperature 0 °C to +55 °C
Electrical connection Terminals max. 2,5 mm
2
, PG11 cable gland
Electrical standard / Case protection EN 61010 / IP22-EN60529
Method of mounting / Ready for working wall mounting / immediately
Material of sample contacting parts Duran glass, PTFE Duran glass, PTFE,
PVDF, Novoprene
Duran glass, PTFE Duran glass, PTFE,
PVDF, Novoprene