14.7
Technical specications and illustrations are without
obligation, subject to modications. 09.96/01.07
Technical Data
205 °C200 °C
B
T
H
( H )
Dimensions
Data required to submit an offer:
Size
Number, position and size of the bushings
Details on built-in components
Flow sheet, assembly drawing
Required inner operating temperature
* Dimensions in ( ) are corresponding to the total height incl. the wall console
Heated sample conditioning housing
Series PAS PAS-100 PAS-200 PAS-300 PAS-400 PAS-500
Assembly interior:
Width mm 320 400 480 400 560
Height mm 240 320 320 400 480
Depth mm 175 250 250 330 400
External dimensions:
Width B mm 470 550 630 550 710
Height H mm 520 (720)* 600 (800)* 600 (800)* 680 (880)* 760 (960)*
Depth T mm 325 400 400 480 550
Weight empty kg 20 28 30 35 50
Ventilator no no no yes yes
Power consumption 600 W 1100 W 1200 W 1400 W 2000 W
Power supply 230V 50/60Hz other voltages on request
Electrical connection 3 m mains cable with an earthing-pin plug; a suitable terminal box is provided for built-in components
Material of cabinet Stainless steel 304
Mounting Wall-mounting with optional wall console
Temperature control Electronically, microprocessor controlled, incl. a solid-state relay
Temperature display Digital LED display
Temperature limit Temperature selection control and limiter in accordance with DIN 12880, protective classification 3.1
Temperature selection Lockable rotary button, Range +20 to +200 °C (Standard)
Operating temperature > 5 °C above ambient temperature up to +200 °C
Ambient temperature max. +40 °C
Case protection IP20 EN 60529
Options: Observation window in door
External temperature control (2 x PT100 internal)
Min. temperature monitoring, status alarm contact, potential free, contact rating 250VAC 16A
EEx explosion-proof heating up to +120 °C inner operating temperature
Steam heating
Wall bushings for sample pipes