M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
info@mc-techgroup.com • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.10
Technical specications and illustrations are without
obligation, subject to modications. 10.04/11.12
Temperature classes for sampling from Ex-zone 0
Options sample tubes
Options Version Part No.
Length 1m, material SS316Ti, temperature max. 600 °C SP2000/SS 20 S 9065
Length 1m, with volume displacer, material SS316Ti, Temperatur max. 600 °C SP2000/SS-Vm 20 S 9067
Length 1m, material Hastelloy-C, temperature max. 900 °C SP2000/HC 20 S 9090
Length 1m, material Kanthal, temperature max. 1300 °C SP2000/KA 20 S 9080
Length 1m, material Incoloy 956, temperature max. 1200 °C SP2000/IC 20 S 9072
Type Possible Options Marking Temperature class Max. process gas temp. (°C) at
the probe entry
Max. surface temperature (°C)
SP3100V
II 1 G / 2 GD
T6 ≤ 68 68
SP3100V
II 1 G / 2 GD
T5 ≤ 80 80
SP3100V
II 1 G / 2 GD
T4 ≤ 108 108
SP3100V /HEX4..
II 1 G / 2 GD
T3 ≤ 160 160
SP3100
II 1 G / 2 GD
T2 ≤ 200 200
Type Possible Options Marking Temperature class Max. process gas temp. (°C) at
the probe entry
Max. surface temperature (°C)
SP3100V
II 1 G / 2 GD
T6 ≤ 80 80
SP3100V
II 1 G / 2 GD
T5 ≤ 95 95
SP3100V /RS, /HEX4-135
II 1 G / 2 GD
T4 ≤ 130 135
SP3100 /HEX4-180
II 1 G / 2 GD
T3 ≤ 195 195
SP3100 /RS, HEX1..
II 2 G / 2 GD
T3 ≤ 195 195
SP3100 /RS
II 2 G / 2 GD
T2 ≤ 200 200
Temperature classes for sampling from Ex-zone 1 or 2
For choosing the adequate sample tube see data sheet 2-1.1.0.6 as well