Gas sampling and gas conditioning technology 3-3.0-ME
8 TECHNICAL DATA
8.1 FOR BASIC COOLER ECM WITHOUT HEAT EXCHANGER
Number of heat exchanger
possibility
range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
at const. conditions: < ±0,1°C
40W at 10°C - 50°C ambient temperature
max. 200VA (start up current at 230V=2,5A; at 120V=4,5A)
240V/120V, 50Hz/60Hz (a):-15%/+10%
terminals: 6-12mm , cable gland: 2xM20x1,5
1 free configurable status alarm with 2 potential free change-over
contacts, alarm point: < 2°C and > 8°C
contact rating 250VAC, 2A; 500 VA; 50W
CAN/CSA-C22.2 No.61010.1-04
UL Std. No. 61010-1(2 Edition)
270 x 270 x 316mm (WxHxD)
12kg/13,5kg (version: 230V/115V)
8.2 OPTIONS FOR BASIC COOLER ECM
Material of heat
exchanger
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
1 pc. incorporated in the cooler, compl. installed, part no.: 01P9125; cooler weight
plus 0,6kg per pump
the sample gas pressure is limited to 2bar absolute
* Standard, other version on request
1
with GL connecting adapter
** Option
*** Maximum values in technical data must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew
point of 5 °C.