6.3
Technical specications and illustrations are without
obligation, subject to modications. 10.99/06.06
* Standard, other version on request.
** Option
*** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and 5 °C outlet dew point.
2)
With GL-Connecting adapter.
3)
With SR25.2 max. 2 bar abs.
GL-adapter and tube ttings for the connection of dierent tube diameters at the heat exchanger see chapter 11, data sheet 11.5 and 11.6.
Technical Data
Order example:
1 cooler ECM-2 with 2 heat exchangers out of glass ECM-2G and 2 peristaltic pumps SR25.2, power 115V 60Hz:
Part. No. 1x 02K7510Xa; 2x 97K0100; 2x 01P9125
Gas cooler series EC
®
version ECM-1 version ECM-2 version ECM-EX2-1 version ECM-EX2-2
Part No. for basis cooler without heat exchanger 02K7500X* (a)** 02K7510X* (a)** 02K7600X* (a)** 02K7610X* (a)**
Number of heat exchanger possibility- 1 2 1 2
Ambient temperature +10 up to +50 °C
Storage temperature -20 to +60°C
Sample outlet dew point range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C
Dew point stability at constant conditions < ±0,1 °C
Sample inlet temperature*** max. 180 °C
Sample inlet dew point*** max. 80 °C
Total cooling capacity 144 KJ/h at +10 up +50 °C ambient
Ready for working <15 min.
Main power connection / power consumption 230V AC* or 115V AC**(a) -15%/+10%, 50/60 Hz, max. 200VA
start up current: 230V 50Hz = 2,5A / 115V 60Hz = 4,5A
Electrical connection terminals 2,5mm
2
, 2x PG11 cable gland (with FM-approval conduit-hub 1/2” NPT)
Status alarm-2 contacts, potential free contact rating 250VAC, 2A, 500VA, 50W, alarm point <+2 °C and >+8 °C*
Case protection / Electrical standard IP20 EN60529 / EN 61010
II 3 G EEx nAC [L] IIC T4 ATEX
Class I, Div.2, Group A, B, C, D, T4 (FM
approval for 115V execution only)
Method of mounting / Case colour wall mounting / RAL 9003
Dimension / Weight 270x270x316 mm (w x h x d) / 12/13,5 kg Version 230V / 115V
Options for basis cooler ECM-1.. and ECM-EX2-1.. ECM-2... and ECM-EX2-2
Heat exchanger type ECM-1G ECM-1PV ECM-1SS ECM-2G ECM-2PV ECM-2SS
Part No. 93K0140 93K0170 93K0160 97K0100 97K0110 97K0115
Material of heat exchanger Duran glass PVDF st. steel 316Ti Duran glass PVDF st. steel 316Ti
Max. gas flow rate per h.e. 250 Nl/hr*** 250 Nl/hr*** 250 Nl/hr*** 150 Nl/hr*** 150 Nl/hr*** 150 Nl/hr***
Gas press. max. bar abs.
3)
2 / 3
2)
3 10* 2 / 3
2)
3 10*
Sample gas connection GL18 for
ø 6 mm
o.d. tube*
G1/4’i G1/4’’i* or
1/4’’NPT**
GL18 for
ø 6 mm
o.d. tube*
Tube ø 6 mm* Tube ø 6 mm
Condensate connection GL⁄25 for
ø 12mm tube*
ø 8mm
or ø 10mm**
G3/8’i G3/8’’i or
3/8 NPT**
GL25 for
ø 12 mm tube*
ø 8 mm
or ø 10 mm**
G3/8’i G3/8’’i or
3/8 NPT**
∆P at max. flow rate 1 mbar 1 mbar 1 mbar 1 mbar 1 mbar 1 mbar
Stagnant space approximately 100 ml 100 ml 100 ml 40 ml 25 ml 30 ml
Peristaltic pump SR25.2 1 pc. incorporated in the cooler, compl. installed, Part No.: 01P9125 cooler weight plus 0,6 kg per pump
Options
Vista de pagina 2