8 Gas sampling and gas conditioning technology 14.1a-ME
6 INTRODUCTION
The gas conditioning CSS-VC unit, either for 19 "- or wall mounting or as a portable unit in carrying
case, is a fully pre-assembled compact continuously-operating gas purification system, which,
depending on the design, can supply a sample gas volume of max. 1 x 250Nl/h (CSS VC1) or 2 x
150Nl/h (CSS-VC2). Sample gas conditioning units are suitable due to their equipment and additional
options for the wide-ranging requirements of continuous gas analysis.
The entire gas conditioning unit is housed in a compact, rugged steel housing so that gas analyses
can be performed rapidly without a great deal of effort, with low maintenance and reliability.
Gas conditioning unit CSS-VC must not be used to support gas/air or gas/oxygen mixtures that are
ignitable during operation, for the supply of combustible gases that can form an ignitable mixture in
combination with air or oxygen, or in potentially explosive atmospheres and in potentially explosive
locations.